Comparison of an automated digital cell morphology analysis system with manual counting
Background Bionovation's CSFA800 is a new automated digital cell imaging analyzer. We evaluated the performance of the CSFA800 by comparing it with artificial peripheral blood white blood cell counting. Methods According to inclusion and exclusion criteria, 131 randomly selected samples (77 abn...
Gespeichert in:
Veröffentlicht in: | iLABMED (Online) 2023-06, Vol.1 (1), p.22-28 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
Bionovation's CSFA800 is a new automated digital cell imaging analyzer. We evaluated the performance of the CSFA800 by comparing it with artificial peripheral blood white blood cell counting.
Methods
According to inclusion and exclusion criteria, 131 randomly selected samples (77 abnormal samples and 54 normal samples) were compared. Correlations between automated and manual counting results were analyzed. Manual counting was carried out according to the guidelines of the Association of Clinical and Laboratory Standards.
Results
Counts of neutrophils, lymphocytes, monocytes, eosinophils, basophils, and immature granulocytes obtained from CSFA800 and artificial methods were linearly and positively correlated, with R values of 0.73, 0.65, 0.24, 0.2, 0.4, and 0.63, respectively, all p |
---|---|
ISSN: | 2834-4448 2834-443X 2834-4448 |
DOI: | 10.1002/ila2.10 |