Experimental Validation of Low-Intensity Beam Extraction Mode for Proton Imaging at the Prometheus Synchrotron
The work is devoted to the dosimetric studies of a low-intensity beam for the implementation of proton radiography on the proton therapy complex Prometeus. In contrast to therapy, radiography using proton beams requires low particle fluxes, less than 1 × 10 6 protons/(s cm 2 ). The controlled unifor...
Gespeichert in:
Veröffentlicht in: | Physics of atomic nuclei 2023-12, Vol.86 (12), p.2732-2737 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The work is devoted to the dosimetric studies of a low-intensity beam for the implementation of proton radiography on the proton therapy complex Prometeus. In contrast to therapy, radiography using proton beams requires low particle fluxes, less than 1 × 10
6
protons/(s cm
2
). The controlled uniform extraction of beams of such intensity at therapeutic accelerators is a significant challenge, and it requires the development of innovative approaches. This study is part of a series of studies on the Russian medical proton synchrotron for implementing the extraction of low-intensity beams. The paper presents data on the key parameter of the extracted beam—the absorbed dose—when performing radiographic plans with a scanning beam. It also provides a quantitative analysis based on computer simulation of the spatial characteristics of proton images obtained using the extracted beam parameters. |
---|---|
ISSN: | 1063-7788 1562-692X |
DOI: | 10.1134/S1063778823110376 |