Determination of persistent organic pollutants by gas chromatography/laser multiphoton ionization/time-of-flight mass spectrometry

An authentic mixture of polychlorinated biphenyls was measured using a short wide-bore capillary column for the group separation of major components present in an actual sample of Kanechlor. The limits of detection were improved by ca. 2 fold in comparison with those obtained using a conventional ca...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Frontiers of environmental science & engineering 2012-02, Vol.6 (1), p.26-31
Hauptverfasser: Shitamichi, Osamu, Matsui, Taiki, Hui, Yamei, Chen, Weiwei, Imasaka, Totaro
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An authentic mixture of polychlorinated biphenyls was measured using a short wide-bore capillary column for the group separation of major components present in an actual sample of Kanechlor. The limits of detection were improved by ca. 2 fold in comparison with those obtained using a conventional capillary column, since the retention time was reduced and the amount of analytes introduced into the mass spectrometer per unit time could be increased. On the other hand, surface-water and sediment samples containing polycyclic aromatic hydrocarbons (PAHs) were collected from the river located in the vicinity of a waste water treatment plant. Even acenaphthylene, a minor component of the mixture could be measured for the sediment sample, and the concentrations were determined for several heavy PAHs. As demonstrated, a technique involving laser multiphoton ionization/time-of-flight mass spectrometry was useful as a sensitive and selective analytical tool for the trace analysis of persistent organic pollutants in an environmental sample.
ISSN:1673-7415
2095-2201
1673-7520
2095-221X
DOI:10.1007/s11783-011-0374-7