Micro-plate magnetic chemiluminescence immunoassay and its applications in carcinoembryonic antigen analysis
A micro-plate magnetic chemiluminescence immunoassay was developed for rapid and high throughput detection of carcinoembryonic antigens (CEA) in human sera. This method was based on a sandwich immunoreaction of fluorescein isothiocyanate (FITC)-labeled anti-CEA antibodies, CEA antigens, and horserad...
Gespeichert in:
Veröffentlicht in: | Science China Chemistry 2010-04, Vol.53 (4), p.812-819 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A micro-plate magnetic chemiluminescence immunoassay was developed for rapid and high throughput detection of carcinoembryonic antigens (CEA) in human sera. This method was based on a sandwich immunoreaction of fluorescein isothiocyanate (FITC)-labeled anti-CEA antibodies, CEA antigens, and horseradish peroxidase (HRP)-conjugated anti-CEA antibodies in micro-plate. The immunomagnetic particles coated with anti-FITC antibodies were used as the solid phase for the immunoassay. The separation procedure was carried out by a magnetic plate adaptor and the luminol-hydrogen peroxide (H
2
O
2
)-HRP system was employed for the chemiluminescence detection. The proposed method combined the advantages of the micro-plate reactor and magnetic particle separation technology with the linear range of 5–250 ng·mL
−1
. The detection limit of CEA was 0.61 ng·mL
−1
. The coefficient of the variation was less than 7% and 13% for intra-assay and inter-assay precision, respectively. Compared with the commercial micro-plate chemiluminescent kit, the proposed method showed a good correlation. |
---|---|
ISSN: | 1674-7291 1862-2771 1869-1870 |
DOI: | 10.1007/s11426-009-0168-y |