New approach to calibrating conductivity meters in the low conductivity range
There is currently a major issue with the calibration of conductivity meters used for high purity water: the lack of availability of a reference material or reference methods for low conductivity ranges (conductivity below 1 μS cm−1 at 25.0 °C, resistivity >1 MΩ cm at 25.0 °C). This paper describ...
Gespeichert in:
Veröffentlicht in: | Accreditation and quality assurance 2005-02, Vol.10 (3), p.78-81 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | There is currently a major issue with the calibration of conductivity meters used for high purity water: the lack of availability of a reference material or reference methods for low conductivity ranges (conductivity below 1 μS cm−1 at 25.0 °C, resistivity >1 MΩ cm at 25.0 °C). This paper describes the current status of conductivity measurements in high purity water. A new and improved approach, currently being investigated, should allow us to make the calibration of conductivity meters used for low conductivity ranges traceable to the SI. |
---|---|
ISSN: | 0949-1775 1432-0517 |
DOI: | 10.1007/s00769-004-0880-4 |