Method for Measuring Noise Power Spectrum Depending on X-ray Tube Angle in Spiral Orbit of Helical CT Scan

Purpose: The noise power spectrum (NPS) in computed tomography (CT) images potentially varies with the X-ray tube angle in a spiral orbit of the helical scan. The purpose of this study was to propose a method for measuring the NPS for each angle of the X-ray tube. Methods: Images of the water phanto...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Japanese Journal of Radiological Technology 2023, Vol.79(10), pp.1144-1150
Hauptverfasser: Ohsugi, Yuki, Narita, Akihiro, Ohkubo, Masaki, Sakai, Kenichi, Fukaya, Takahiro, Noto, Yoshiyuki
Format: Artikel
Sprache:jpn
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose: The noise power spectrum (NPS) in computed tomography (CT) images potentially varies with the X-ray tube angle in a spiral orbit of the helical scan. The purpose of this study was to propose a method for measuring the NPS for each angle of the X-ray tube. Methods: Images of the water phantom were acquired using a helical scan. As a conventional method, we measured the two-dimensional (2D) NPS from each image and averaged them; the obtained 2D-NPS was referred to as NPSconventional. In the proposed method, we made the X-ray tube angle θ (0°≤θ0.999) between the rotation angle of NPSθ and the θ. Conclusion: The proposed method was demonstrated to be effective for evaluating anisotropic noise characteristics depending on the X-ray tube angle.