Revision total hip arthroplasty with cemented femoral component
We reviewed retrospectively the results of 28 hips (25 patients) after revision of the femoral component with use of a cemented stem, because of aseptic loosening. The mean duration of follow-up was 4.43 years (range 2–12 years). Over the course of the study period, repeat revision was done in 4 hip...
Gespeichert in:
Veröffentlicht in: | European journal of orthopaedic surgery & traumatology 2008-07, Vol.18 (5), p.327-332 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We reviewed retrospectively the results of 28 hips (25 patients) after revision of the femoral component with use of a cemented stem, because of aseptic loosening. The mean duration of follow-up was 4.43 years (range 2–12 years). Over the course of the study period, repeat revision was done in 4 hips after an average of 4.45 years. Three hips had a repeat revision of the femoral component because of aseptic loosening and one for a deep infection. The rate of loosening of the femoral component was 32.4% (9 hips) at an average of 5.22 years. The 5-year survival rate was 76.9% with mechanical failure as end point; and 90% with re-revision of femoral component because of aseptic loosening as end point. The cement mantle was the principal factor, which was significantly associated with a better survival rate of femur fixation (
P
|
---|---|
ISSN: | 1633-8065 1432-1068 |
DOI: | 10.1007/s00590-008-0310-0 |