Estimation of the Thickness Profile of a Human Skull Phantom by Ultrasound Methods Using a Two-Dimensional Array

The paper presents the results of evaluating the thickness profile of a skull phantom using a two-dimensional ultrasound array consisting of piezoelectric elements with a center frequency of 2.1 MHz. Two pulse-echo ultrasound methods were used in the experiment: the A-mode elementwise measurements a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Acoustical physics 2023-02, Vol.69 (1), p.112-118
Hauptverfasser: Asfandiyarov, S. A., Rosnitskiy, P. B., Tsysar, S. A., Yuldashev, P. V., Khokhlova, V. A., Sinitsyn, V. E., Mershina, E. A., Sapozhnikov, O. A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The paper presents the results of evaluating the thickness profile of a skull phantom using a two-dimensional ultrasound array consisting of piezoelectric elements with a center frequency of 2.1 MHz. Two pulse-echo ultrasound methods were used in the experiment: the A-mode elementwise measurements and scanning with a focused probing beam created by the entire array using delay-and-sum (DAS) beamforming. The obtained thickness profiles are compared with the reference thickness profile obtained using X-ray computed tomography. It was shown that ultrasound DAS beamforming with a focused probing beam makes it technically possible to estimate the thickness profile of the skull phantom.
ISSN:1063-7710
1562-6865
DOI:10.1134/S106377102270004X