Estimation of the Thickness Profile of a Human Skull Phantom by Ultrasound Methods Using a Two-Dimensional Array
The paper presents the results of evaluating the thickness profile of a skull phantom using a two-dimensional ultrasound array consisting of piezoelectric elements with a center frequency of 2.1 MHz. Two pulse-echo ultrasound methods were used in the experiment: the A-mode elementwise measurements a...
Gespeichert in:
Veröffentlicht in: | Acoustical physics 2023-02, Vol.69 (1), p.112-118 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The paper presents the results of evaluating the thickness profile of a skull phantom using a two-dimensional ultrasound array consisting of piezoelectric elements with a center frequency of 2.1 MHz. Two pulse-echo ultrasound methods were used in the experiment: the A-mode elementwise measurements and scanning with a focused probing beam created by the entire array using delay-and-sum (DAS) beamforming. The obtained thickness profiles are compared with the reference thickness profile obtained using X-ray computed tomography. It was shown that ultrasound DAS beamforming with a focused probing beam makes it technically possible to estimate the thickness profile of the skull phantom. |
---|---|
ISSN: | 1063-7710 1562-6865 |
DOI: | 10.1134/S106377102270004X |