Imaging Capabilities of the IRIS and IRIS XL-260 PET/CT Systems for High-Throughput Imaging: a Quantification Cross Validation Study

The concept of imaging several subjects simultaneously is an active preclinical research topic. In the present paper, we assessed the imaging capabilities of two PET/CT systems based on similar detector technology. The IRIS system is a mice/rat imaging system arranged in two rings of eight detectors...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on radiation and plasma medical sciences 2023-04, Vol.7 (4), p.1-1
Hauptverfasser: Boisson, F, Hourlier, A, Bekaert, V, Thomas, L, Marchand, P, Santiago, E, Cao, L, Brasse, D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The concept of imaging several subjects simultaneously is an active preclinical research topic. In the present paper, we assessed the imaging capabilities of two PET/CT systems based on similar detector technology. The IRIS system is a mice/rat imaging system arranged in two rings of eight detectors each, the more recent IRIS XL 260, dedicated to Non-Human Primate (NHP) imaging has a single ring of sixteen detectors. Both systems were equipped with standard animal cells (Minerve) and 3-mice adapters. Our objective is to study which of these systems could be more appropriate to perform quantitative high-throughput imaging on mice. Phantoms and 18FFDG tumor-bearing mice acquisitions have been conducted in single and high-throughput modes using both the IRIS (up to 3 mice) and the IRIS XL-260 PET/CT (up to 6 mice) systems. Image Quality Phantom results obtained in high-throughput mode show some slight degradation of the recovery coefficient for rods of 1, 2 and 3 mm in diameter compared to the single-mode results, as one would normally expect. Similarly, we observed a decrease in image uniformity between the single-and the high-throughput modes for both the IRIS PET and the IRIS XL-260 PET systems. We performed 18FFDG tumor-bearing mice PET acquisitions with both systems. In order to estimate the quantification differences in all the studied configurations, we calculate the ratios between the %ID/g values extracted from the PET images and the ex-vivo values. In single mode acquisitions, ratios of 0.94 ± 0.09 and 0.83 ± 0.08 were obtained for the IRIS and the IRIS XL-260 PET systems, respectively. In high-throughput mode, ratios of 0.78 ± 0.12 and 0.73 ± 0.13 were obtained for the IRIS and the IRIS XL-260 PET systems, respectively. The difference in %ID/g between static acquisitions and ex-vivo value is not statistically significant (p-value > 0.1) for both PET systems in single mode, as well as for the IRIS system in high-throughput mode. We noted a slightly higher statistical difference between static acquisitions and ex-vivo values for the IRIS XL-260 system with a p-value of 0.015. Phantoms and in-vivo studies have made it possible to highlight the capability of the two systems to perform high-throughput acquisitions. Our results suggest that the IRIS system configuration may be the most suitable when aiming for quantitative high-throughput mice imaging. The large transverse field of view of the IRIS XL-260 makes it possible to image a greater number of mi
ISSN:2469-7311
2469-7303
DOI:10.1109/TRPMS.2022.3222565