Studies on the radiation dose, image quality and low contrast detectability from MSCT abdomen by using low tube voltage
Background To study radiation dose, image quality and low-contrast cylinder detectability from multislice CT (MSCT) abdomen by using low tube voltage using the American College of Radiology (ACR) phantom. The ACR phantom (low-contrast module) was scanned with 64 MSCT scanner (Brilliance, Philips Med...
Gespeichert in:
Veröffentlicht in: | Egyptian Journal of Radiology and Nuclear Medicine 2021-11, Vol.52 (1), p.269-7, Article 269 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
To study radiation dose, image quality and low-contrast cylinder detectability from multislice CT (MSCT) abdomen by using low tube voltage using the American College of Radiology (ACR) phantom. The ACR phantom (low-contrast module) was scanned with 64 MSCT scanner (Brilliance, Philips Medical System, Eindhoven, Netherlands) with 80 and 120 KVP, utilizing different tube current time product (mAs) range from 50 to 380 mAs. The image noise (SD), signal to noise ratio, contrast-to-noise ratio (CNR), and scores of low contrast detectability were assessed for every image respectively.
Results
From images analyses, the noise essentially increased with the use of low tube voltage. The CNR was 0.94 ± 0.27 at 120 KVP, and CNR was 0.43 ± 0.22 at 80 KVP. However, with the same dose, there were no differences of statistical significance in scores of low-contrast detectability between 120 KVP at 300mAs and 80 KVP at (200–380) mAs (
p
> 0.05). At 300 mAs, the CTDI
vol
obtained at 80 KVP was about 29% of that at 120 KVP. The CTDI
vol
obtained at 80 KVP were decreased from 5% at 50 mAs, to 37% at 380 mAs.
Conclusions
There is a possibility to decrease exposure of radiation virtually by reducing KVP from 120 to 80 KVP in examination of abdominal CT when the high tube current is used, though increasing image noise at low tube voltage. |
---|---|
ISSN: | 2090-4762 0378-603X 2090-4762 |
DOI: | 10.1186/s43055-021-00613-y |