Practical realization and design of three-cascade prism mass analyzer for isotopic analysis of H-He mixtures using MS-platform technology
The paper presents the construction of a mass analyzer for the analysis of hydrogen-helium mixtures based on a magnetic prism using a three-stage circuit that includes two cylindrical capacitors. The choice of an ion-optical scheme that provides a high resolution of 3300, which is necessary for the...
Gespeichert in:
Veröffentlicht in: | Journal of physics. Conference series 2022-12, Vol.2388 (1), p.12020 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The paper presents the construction of a mass analyzer for the analysis of hydrogen-helium mixtures based on a magnetic prism using a three-stage circuit that includes two cylindrical capacitors. The choice of an ion-optical scheme that provides a high resolution of 3300, which is necessary for the separation of all nuclides formed during the analysis of hydrogen-helium mixtures, is justified. It is shown that the innovative «MS-Platform» technology makes it possible to ensure high assembly accuracy of the elements of the prism mass analyzer ion-optical circuit, which is necessary to realize all the advantages of a prism mass spectrometer compared to sector ones. |
---|---|
ISSN: | 1742-6588 1742-6596 |
DOI: | 10.1088/1742-6596/2388/1/012020 |