Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the KRISS, Republic of Korea, and the BIPM in low-energy x-rays

A key comparison has been made between the air-kerma standards of the KRISS, Republic of Korea, and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the standard uncertainty of the comparison of 2.2 parts in 10 3 . The results are analysed and...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Metrologia 2022-01, Vol.59 (1A), p.6011
Hauptverfasser: Burns, D T, Kessler, C, Yi, Chul-Young, Ho Kim, Yun
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A key comparison has been made between the air-kerma standards of the KRISS, Republic of Korea, and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the standard uncertainty of the comparison of 2.2 parts in 10 3 . The results are analysed and presented in terms of degrees of equivalence, suitable for entry into the BIPM key comparison database. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/ . The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
ISSN:0026-1394
1681-7575
DOI:10.1088/0026-1394/59/1A/06011