RSC: Dosimetry in high-dose-rate brachytherapy with a radio-fluorogenic gel dosimeter

A nanoclay-based radio-fluorogenic gel (NC-RFG) was used to verify the source position and dose distribution in high-dose-rate (HDR) brachytherapy. The dose response confirmed linearity up to 60 Gy. The source position could be detected with an accuracy of ≤0.3 mm, and the dose distribution near the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of physics. Conference series 2022-01, Vol.2167 (1), p.12032
Hauptverfasser: Inoue, K., Watanabe, Y., Maeyama, T., Mizukami, S., Hayashi, S., Terazaki, T., Muraishi, H., Gomi, T., Shimono, T.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A nanoclay-based radio-fluorogenic gel (NC-RFG) was used to verify the source position and dose distribution in high-dose-rate (HDR) brachytherapy. The dose response confirmed linearity up to 60 Gy. The source position could be detected with an accuracy of ≤0.3 mm, and the dose distribution near the Ir-192 source showed good agreement with the Monte Carlo simulation. NC-RFG can be expected to be a quality assurance tool suitable for the evaluating the dose distribution in HDR brachytherapy.
ISSN:1742-6588
1742-6596
DOI:10.1088/1742-6596/2167/1/012032