Comparison of the X-ray tube spectrum measurement using BGO, NaI, LYSO, and HPGe detectors in a preclinical mini-CT scanner: Monte Carlo simulation and practical experiment

In diagnostic X-ray computed tomography (CT) imaging, some applications, such as dose measurement using the Monte Carlo method and material decomposition using dual/multi-energy approaches, rely on accurate knowledge of the energy spectrum of the X-ray beam. In this regard, X-ray detectors providing...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiation physics and chemistry (Oxford, England : 1993) England : 1993), 2021-12, Vol.189, p.109666, Article 109666
Hauptverfasser: Lohrabian, Vahid, Kamali-Asl, Alireza, Harvani, Hossein Ghadiri, Hosseini Aghdam, Seyed Rashid, Arabi, Hossein, Zaidi, Habib
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In diagnostic X-ray computed tomography (CT) imaging, some applications, such as dose measurement using the Monte Carlo method and material decomposition using dual/multi-energy approaches, rely on accurate knowledge of the energy spectrum of the X-ray beam. In this regard, X-ray detectors providing an accurate estimation of the X-ray spectrum could greatly impact the quality of dual/multi-energy CT imaging and patient-specific dosimetry. The aim of this study is to estimate the intrinsic efficiency and energy resolution of different types of solid-state gamma-ray detectors in order to generate a precise dual-energy X-ray beam from the conventional x-ray tube using external X-ray filters. The X-ray spectrum of a clinical X-ray tube was experimentally measured using a high purity Germanium detector (HPGe) and the obtained spectrum validated by Monte Carlo (MC) simulations. The obtained X-ray spectrum from the experiment was employed to assess the energy resolution and detection efficiency of different inorganic scintillators and semiconductor-based solid-state detectors, namely HPGe, BGO, NaI, and LYSO, using MC simulations. The best performing detector was employed to experimentally create and measure a dual-energy X-ray spectrum through applying attenuating filters to the original X-ray beam. The simulation results indicated 9.16% energy resolution for the HPGe detector wherein the full width-at-half-maximum (FWHM) of the energy resolution for the HPGe detector was about 1/3rd of the other inorganic detectors. The X-ray spectra estimated from the various source energies exhibited a good agreement between experimental and simulation results with a maximum difference of 6%. Owing to the high-energy discrimination power of the HPGe detector, a dual-energy X-ray spectrum was created and measured from the original X-ray spectrum using 0.5 and 4.5 mm Aluminum external filters, which involves 70 and 140 keV energy peaks with 8% overlap. The experimental measurements and MC simulations of the HPGe detector exhibited close agreement in high-energy resolution estimation of the X-ray spectrum. Given the accurate measurement of the X-ray spectrum with the HPGe detector, a dual-energy X-ray spectrum was generated with minimal energy overlap using external X-ray filters. •Accurate knowledge of the energy spectrum of the x-ray beam.•Estimate the intrinsic efficiency and energy resolution of different types detectors.•Assess the energy resolution and detection efficiency
ISSN:0969-806X
1879-0895
DOI:10.1016/j.radphyschem.2021.109666