Dosimetric performances of optically detected Fricke-agarose-Xylenol Orange gel
A first prototype of optical scanner based on a CCD camera and the use of Fricke-agarose-XO has allowed us to reconstruct 3D dose maps, monitoring doses in the range 0.5–3 Gy. Measurements performed with laboratory tests and with clinical beams would indicate that the uncertainty involved in dose ma...
Gespeichert in:
Veröffentlicht in: | Journal of physics. Conference series 2009-05, Vol.164 (1), p.012015 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A first prototype of optical scanner based on a CCD camera and the use of Fricke-agarose-XO has allowed us to reconstruct 3D dose maps, monitoring doses in the range 0.5–3 Gy. Measurements performed with laboratory tests and with clinical beams would indicate that the uncertainty involved in dose mapping are compatible with requirements of a 3D dosimetry for the verifications of treatment plans. |
---|---|
ISSN: | 1742-6596 1742-6588 1742-6596 |
DOI: | 10.1088/1742-6596/164/1/012015 |