Design of clinical trials to assess diabetes treatment: Minimum duration of continuous glucose monitoring data to estimate time‐in‐ranges with the desired precision
ABSTRACT Aim To compute the uncertainty of time‐in‐ranges, such as time in range (TIR), time in tight range (TITR), time below range (TBR) and time above range (TAR), to evaluate glucose control and to determine the minimum duration of a trial to achieve the desired precision. Materials and Methods...
Gespeichert in:
Veröffentlicht in: | Diabetes, obesity & metabolism obesity & metabolism, 2021-11, Vol.23 (11), p.2446-2454 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | ABSTRACT
Aim
To compute the uncertainty of time‐in‐ranges, such as time in range (TIR), time in tight range (TITR), time below range (TBR) and time above range (TAR), to evaluate glucose control and to determine the minimum duration of a trial to achieve the desired precision.
Materials and Methods
Four formulas for the aforementioned time‐in‐ranges were obtained by estimating the equation's parameters on a training set extracted from study A (226 subjects, ~180 days, 5‐minute Dexcom G4 Platinum sensor). The formulas were then validated on the remaining data. We also illustrate how to adjust the parameters for sensors with different sampling rates. Finally, we used study B (45 subjects, ~365 days, 15‐minute Abbott Freestyle Libre sensor) to further validate our results.
Results
Our approach was effective in predicting the uncertainty when time‐in‐ranges are estimated using n days of continuous glucose monitoring (CGM), matching the variability observed in the data. As an example, monitoring a population with TIR = 70%, TITR = 50%, TBR = 5% and TAR = 25% for 30 days warrants a precision of ±3.50%, ±3.68%, ±1.33% and ±3.66%, respectively.
Conclusions
The presented approach can be used to both compute the uncertainty of time‐in‐ranges and determine the minimum duration of a trial to achieve the desired precision. An online tool to facilitate its implementation is made freely available to the clinical investigator. |
---|---|
ISSN: | 1462-8902 1463-1326 |
DOI: | 10.1111/dom.14483 |