Phase calibration of hydrophones: heterodyne time-delay spectrometry and broadband pulse technique using an optical reference hydrophone

Two hydrophone calibration techniques are presented which provide amplitude and phase of hydrophone sensitivity. Heterodyne time-delay spectrometry can be applied in the frequency range from 0.5 to 50 MHz to all common hydrophones but needs a standard to obtain absolute values. A pulse technique bet...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of physics. Conference series 2004-01, Vol.1 (1), p.14-19
Hauptverfasser: Koch, Ch, Wilkens, V
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Two hydrophone calibration techniques are presented which provide amplitude and phase of hydrophone sensitivity. Heterodyne time-delay spectrometry can be applied in the frequency range from 0.5 to 50 MHz to all common hydrophones but needs a standard to obtain absolute values. A pulse technique between 1 and 70 MHz allows exploiting of an optical multilayer hydrophone as a phase standard. In combination with HTDS, a calibration service is established that covers the complete spectrum of common piezoelectric and optical hydrophones for absolute amplitude and phase calibration. The measured complex frequency responses form the basis of deconvolution procedures suitable for the correction of measurements of, for example, the output of ultrasound diagnostic machines.
ISSN:1742-6596
1742-6588
1742-6596
DOI:10.1088/1742-6596/1/1/006