The Australian MRI-Linac Program: measuring profiles and PDD in a horizontal beam
The Australian MRI-Linac consists of a fixed horizontal photon beam combined with a MRI. Commissioning required PDD and profiles measured in a horizontal set-up using a combination of water tank measurements and gafchromic film. To validate the methodology, measurements were performed comparing PDD...
Gespeichert in:
Veröffentlicht in: | Journal of physics. Conference series 2017-02, Vol.777 (1), p.12035 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The Australian MRI-Linac consists of a fixed horizontal photon beam combined with a MRI. Commissioning required PDD and profiles measured in a horizontal set-up using a combination of water tank measurements and gafchromic film. To validate the methodology, measurements were performed comparing PDD and profiles measured with the gantry angle set to 0 and 90° on a conventional linac. Results showed agreement to within 2.0% for PDD measured using both film and the water tank at gantry 90° relative to PDD acquired using gantry 0°. Profiles acquired using a water tank at both gantry 0 and 90° showed agreement in FWHM to within 1 mm. The agreement for both PDD and profiles measured at gantry 90° relative to gantry 0° curves indicates that the methodology described can be used to acquire the necessary beam data for horizontal beam lines and in particular, commissioning the Australian MRI-linac. |
---|---|
ISSN: | 1742-6588 1742-6596 |
DOI: | 10.1088/1742-6596/777/1/012035 |