The Energy Characteristics of a Multichannel Scintillation Spectrometer
— A multichannel scintillation spectrometer with a thickness of 2.5X 0 has been calibrated on a secondary electron beam with energies of 23−280 MeV at the Pakhra accelerator of the Lebedev Physical Institute. The relative energy resolution of the spectrometer is determined by the electron energy and...
Gespeichert in:
Veröffentlicht in: | Instruments and experimental techniques (New York) 2021-09, Vol.64 (5), p.663-668 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | —
A multichannel scintillation spectrometer with a thickness of 2.5X
0
has been calibrated on a secondary electron beam with energies of 23−280 MeV at the Pakhra accelerator of the Lebedev Physical Institute. The relative energy resolution of the spectrometer is determined by the electron energy and the spectrometer thickness. The best relative energy resolution of the spectrometer is attained at an electron energy of ~90 MeV; its values are δ = 13 and 9.4% at spectrometer thicknesses of 1.0X
0
and 2.5X
0
, respectively. |
---|---|
ISSN: | 0020-4412 1608-3180 |
DOI: | 10.1134/S0020441221050018 |