Development of methods for safe MR-scanning of patients with implantable medical devices. Metal structures heating

The paper presents the experience of experimental determination of implanted medical devices heating value during an MR-study. This work is a part of the development of a comprehensive methodology for ensuring the safety of patients with metal structures. Objects of various shapes and materials were...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of physics. Conference series 2020-03, Vol.1461 (1), p.12152
Hauptverfasser: Semenov, D S, Vasilev, Yu A, Akhmad, E S
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The paper presents the experience of experimental determination of implanted medical devices heating value during an MR-study. This work is a part of the development of a comprehensive methodology for ensuring the safety of patients with metal structures. Objects of various shapes and materials were placed in a gel phantom, scanning was performed on an MR scanner with a static magnetic field of 1.5 T. It is shown that the temperature change in close proximity to the surface of a metal object can be within the permissible normative documentation limits or be negligible. The existing in Russia indications for magnetic resonance imaging (MRI) for patients with various metal objects should be reviewed.
ISSN:1742-6588
1742-6596
DOI:10.1088/1742-6596/1461/1/012152