A Comparative Study of Hyperelastic Constitutive Models to Characterize the Behaviour of a Biopolymer Material for Diaphragm of Blood Pump Manufacturing
This article presents the results of the stress-strain state modelling of a blood pump diaphragm manufacturing from polyurethane 'Vitur' T-1413-85 for pulsatile type Ventricle Assist Device systems. Based on the analysis of the most distributed models of hyperelasticity, a model which most...
Gespeichert in:
Veröffentlicht in: | IOP conference series. Materials Science and Engineering 2020-01, Vol.739 (1), p.12041 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This article presents the results of the stress-strain state modelling of a blood pump diaphragm manufacturing from polyurethane 'Vitur' T-1413-85 for pulsatile type Ventricle Assist Device systems. Based on the analysis of the most distributed models of hyperelasticity, a model which most accurately describes the behaviour of the specified material has been determined (R2 = 0.9999 between numerical simulation and experimental uniaxial tensile tests data). Experimental studies have been carried out to determine the displacement of the diaphragm with a thickness of 0.3 mm and 0.5mm under the influence of control pressure. A comparison of the modelling results of the blood pump diaphragm stress-strain state and the conducted experimental studies on the determination of displacements confirmed the adequacy of the chosen model (R2 = 0.96 between numerical simulation and experiment) for both of the studied diaphragm thickness. |
---|---|
ISSN: | 1757-8981 1757-899X |
DOI: | 10.1088/1757-899X/739/1/012041 |