Comparison of intrapelvic pressures during flexible ureteroscopy, mini-percutaneous nephrolithotomy, standard percutaneous nephrolithotomy, and endoscopic combined intrarenal surgery in a kidney model
Purpose To compare intrapelvic pressure (IPP) levels achieved during f-URS, mini-PCNL, standard PCNL, and endoscopic combined intrarenal surgery in a kidney model. Methods A silicone model simulating the complete urinary tract was used for all the experiments. We compared: a 9.5Fr f-URS, a 12Fr mini...
Gespeichert in:
Veröffentlicht in: | World journal of urology 2021-07, Vol.39 (7), p.2709-2717 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To compare intrapelvic pressure (IPP) levels achieved during f-URS, mini-PCNL, standard PCNL, and endoscopic combined intrarenal surgery in a kidney model.
Methods
A silicone model simulating the complete urinary tract was used for all the experiments. We compared: a 9.5Fr f-URS, a 12Fr mini-nephroscope and a 26Fr nephroscope. The irrigation pressure was set at 40 and 193 cmH
2
O. We compared: f-URS-S ± ureteral access sheath (UAS, 10/12Fr, 11/13Fr, 12/14Fr) ± 273 μm laser fiber, Mini-PCNL with different sizes of operating sheath (15/16Fr, 16.5/17.5Fr, 21/22Fr) ± 365 μm laser fiber, Standard PCNL with an operating sheath of 30Fr ± Lithotripter LithoClast Master 11.4Fr.
Results
f-URS: IPP values ranged between 1.4 and 46.2 cmH
2
O. Factors reducing IPP were an irrigation pressure at 40 cmH
2
O, an occupied working channel, and the use of a UAS except with the 10/12Fr at 193 cmH
2
O. Mini-PCNL: IPP values ranged between 2.4 and 39.7 cmH
2
O. Factors reducing IPP were irrigation pressure at 40 cmH
2
O, a large operating sheath (> 15/16Fr). The occupation of the working channel did not affect the IPP at 40 cmH
2
O, while it decreased at 193 cmH
2
O.
Standard PCNL: IPP values ranged between 1.4 and 7.3 cmH
2
O. Occupancy of the working channel did not affect IPP at 40 cmH
2
O, while it increased at 193 cmH
2
O.
Conclusion
We recorded for the first time IPP values according to different endourological techniques and configurations. IPP never exceed 50 cmH
2
O irrespectively of the assessed technique/setup. The factors reducing IPP were a low irrigation pressure (40 cmH
2
O), the use of a UAS or a working sheath appropriate to the diameter of the endoscope, as well as the occupation of the working channel in the case of f-URS. |
---|---|
ISSN: | 0724-4983 1433-8726 |
DOI: | 10.1007/s00345-020-03450-2 |