Enzyme-free colorimetric determination of uric acid based on inhibition of gold nanorods etching

•An enzyme-free colorimetric method for sensitive detection of UA was established.•LSPR absorption peak shift (Δλ) of AuNRs before and after inhibition was used to quantitative determination of UA.•The proposed method could overcome the limitation of deviation from beer's law caused by too low...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Sensors and actuators. B, Chemical Chemical, 2021-04, Vol.333, p.129638, Article 129638
Hauptverfasser: Qin, Xiu, Yuan, Chunling, Geng, Guoxing, Shi, Rui, Cheng, Shiqi, Wang, Yilin
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:•An enzyme-free colorimetric method for sensitive detection of UA was established.•LSPR absorption peak shift (Δλ) of AuNRs before and after inhibition was used to quantitative determination of UA.•The proposed method could overcome the limitation of deviation from beer's law caused by too low or too high absorbance. An enzyme-free colorimetric method for sensitive detection of uric acid (UA) was established, which was achieved by inhibiting the etching of gold nanorods (AuNRs). The longitudinal direction of AuNRs would be gradually etched by I2, as an oxidant, generated by the reaction between MnO2 nanosheets (MnO2 NS) and I−. Benefitted from the length change, the localized surface plasmon resonance (LSPR) peak of AuNRs showed an obvious blue shift, led to the color change of AuNRs solution. Once UA was introduced, MnO2 NS was reduced to Mn2+ rapidly. As a result, the formation of I2 was reduced and the etching of AuNRs was inhibited. Therefore, a simple method for the assay of UA was achieved with UV–vis absorption spectra and bare eyes, respectively. In optimization conditions, LSPR absorption peak shift (Δλ) of AuNRs before and after inhibition enabled the sensitive detection of UA in the range of 0.8−30 μM and 30−300 μM, the detection limit of low concentration range was 0.76 μM. The developed method has been utilized for determining UA in clinical samples with satisfactory results.
ISSN:0925-4005
1873-3077
DOI:10.1016/j.snb.2021.129638