Development of the CMC for air kerma Ka for x-ray reference beams in the RQR-M qualities of the IEC 61267:2005 at the SSDL-ININ
The SSDL-ININ developed the Calibration Measurement Capabilities (CMC) for air kerma (Ka) for X-ray reference beams in the IEC 61267:2005 RQR-M qualities for mammography. In this work the following parameters were determined: namely, the practical peak voltage (PPV) and the first half-value layer (H...
Gespeichert in:
Veröffentlicht in: | AIP conference proceedings 2021-04, Vol.2348 (1) |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The SSDL-ININ developed the Calibration Measurement Capabilities (CMC) for air kerma (Ka) for X-ray reference beams in the IEC 61267:2005 RQR-M qualities for mammography. In this work the following parameters were determined: namely, the practical peak voltage (PPV) and the first half-value layer (HVL1). Additionally, Monte Carlo (MC) simulations are done to calculate their mean energies (E¯), and to verify that these parameters’ values are within the tolerances indicated by the IEC 61267:2005. |
---|---|
ISSN: | 0094-243X 1551-7616 |
DOI: | 10.1063/5.0051643 |