A Simple and High‐Accuracy PID‐Based Temperature Control System for Magnetic Hyperthermia Using Fiber Optic Thermometer

Magnetic hyperthermia is very promising cancer therapy in which heating temperature is expected to be a key determinant for its success. Herein, we present a simple, real‐time, and high‐accuracy proportional‐integral‐derivative‐based temperature control system for magnetic hyperthermia using widely‐...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEJ transactions on electrical and electronic engineering 2021-05, Vol.16 (5), p.807-809
Hauptverfasser: Shikano, Akihiro, Tonthat, Loi, Yabukami, Shin
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Magnetic hyperthermia is very promising cancer therapy in which heating temperature is expected to be a key determinant for its success. Herein, we present a simple, real‐time, and high‐accuracy proportional‐integral‐derivative‐based temperature control system for magnetic hyperthermia using widely‐used induction heater and fiber optic thermometer. Our in vitro result shows that the temperature of clinically‐approved Resovist® (0.5 ml) was accurately controlled with very small overshoot (e.g., 45.0 ± 0.07 °C and 55.0 ± 0.11°C for low‐ and high‐temperature magnetic hyperthermia). Our system has the potential to be used for magnetic hyperthermia either alone or together with other cancer treatment modalities for basic and clinical studies. © 2021 Institute of Electrical Engineers of Japan. Published by Wiley Periodicals LLC.
ISSN:1931-4973
1931-4981