Comparison between three weighing methods for source preparation in radionuclide metrology
In radionuclide metrology three weighing methods could be used for source preparation. The Pycnometer, Elimination and Modified Elimination methods should be compared to evaluate the compatibility in the deposited mass results, so increasing their reliability. In this work, a sequence of micro drop...
Gespeichert in:
Veröffentlicht in: | Journal of physics. Conference series 2021-03, Vol.1826 (1), p.12038 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | In radionuclide metrology three weighing methods could be used for source preparation. The Pycnometer, Elimination and Modified Elimination methods should be compared to evaluate the compatibility in the deposited mass results, so increasing their reliability. In this work, a sequence of micro drop deposition weighing is presented which allows comparing the results by the same drop deposition. The results from each method were determined and their compatibility was checked. As a result, the three methods based on carefully weighing were found compatible and comply with the required threshold uncertainty. Furthermore, it was confirmed the results are traceable to mass the mass unit. |
---|---|
ISSN: | 1742-6588 1742-6596 |