Elective replacement of a 4‐year‐old leadless pacemaker: New approach for end‐of‐life management

We report the first case of new technique of replacement of a Micra TPS, due to battery depletion. A 38‐year‐old patient was admitted due to battery depletion of a TPS, after 44 months of regular pacemaker functioning. After routine implantation of a new TPS, we use a snare loop inserted in the deli...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of cardiovascular electrophysiology 2021-04, Vol.32 (4), p.1171-1173
Hauptverfasser: De Filippo, Paolo, Malanchini, Giovanni, Leidi, Cristina, Racheli, Marco, Ferrari, Gabriele, Senni, Michele, Ferrari, Paola
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report the first case of new technique of replacement of a Micra TPS, due to battery depletion. A 38‐year‐old patient was admitted due to battery depletion of a TPS, after 44 months of regular pacemaker functioning. After routine implantation of a new TPS, we use a snare loop inserted in the delivery system to capture the old TPS. We believe this approach a good option not to abandon the depleted device, to avoid possible electrical interference or space occupation in right ventricle. This new approach allows to change the strategy during procedure and does not increase significantly the procedure costs.
ISSN:1045-3873
1540-8167
DOI:10.1111/jce.14946