0445 A Novel Algorithm for the Estimation of Sleep States Based on Breathing and Movement
Abstract Introduction We tested the diagnostic accuracy of the novel Nox BodySleep™ algorithm (Nox Medical, Iceland) for the estimation of sleep states from polygraphy (PG) sleep recordings based on features extracted from actigraphy and respiratory inductance plethysmography (RIP) belts. The algori...
Gespeichert in:
Veröffentlicht in: | Sleep (New York, N.Y.) N.Y.), 2020-05, Vol.43 (Supplement_1), p.A170-A170 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Introduction
We tested the diagnostic accuracy of the novel Nox BodySleep™ algorithm (Nox Medical, Iceland) for the estimation of sleep states from polygraphy (PG) sleep recordings based on features extracted from actigraphy and respiratory inductance plethysmography (RIP) belts. The algorithm automatically classifies epochs into three states, Wake, REM sleep and NonREM sleep. Validation was performed against polysomnography (PSG) in a sleep laboratory collective including patients with sleep disordered breathing (SBAS) and sleep related movements disorders.
Methods
Patients received PSG according to clinical routine. The recording was evaluated by the novel algorithm and the results were evaluated by descriptive statistics methods (IBM SPSS Statistics 25.0).
Results
We found a good Spearman correlation (r=0.8) and a bias of 11 minutes for the estimation of Total Sleep Time. Sleep Efficiency was also valued with a good Spearman correlation (r=0.7) and a bias of 1.6%. Wake phases were estimated with a F1 score of 0.64 while REM and Non-REM phases were evaluated with a F1 score of 0.73 and 0.82, respectively. Additionally, an overall accuracy of 0.8 and a Cohens kappa of 0.7 were found. Patients with sleep related movement disorders showed a slighly weaker correlation as patients with SBAS.
Conclusion
The algorithm shows a good diagnostic accuracy for the estimation of sleep states and significant sleep parameters. After validation on a larger patient collective, it could be used in the ambulatory and telemedical field to allow investigations comparable to the accuracy of a PSG.
Support
No support. |
---|---|
ISSN: | 0161-8105 1550-9109 |
DOI: | 10.1093/sleep/zsaa056.442 |