Verification of a Capacitive Voltage Divider With 6-μrad Uncertainty Up To 100 kV

A reference setup for system calibration of industrial transformer loss measurement (TLM) systems consists of three main components, a voltage divider, a current transformer, and a power meter, and their phase displacements should each not be more than 10 μrad to achieve an overall system uncertaint...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on instrumentation and measurement 2021, Vol.70, p.1-9, Article 1004809
Hauptverfasser: Ye, Gu, Zhao, Wei, Rietveld, Gert
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A reference setup for system calibration of industrial transformer loss measurement (TLM) systems consists of three main components, a voltage divider, a current transformer, and a power meter, and their phase displacements should each not be more than 10 μrad to achieve an overall system uncertainty of better than 20 μW/VA. We have extensively verified the uncertainty level of the current-comparator-based capacitive voltage divider (CVD) used in the TLM reference setup of the Van Swinden Laboratorium (VSL), both from the component level and the system as a whole. Different practical conditions relevant for on-site measurements are considered, e.g., measurement cable lengths, cable types, and grounding. The verification measurement results show an agreement of better than (6 ± 6) × 10 -6 in ratio error and (4 ± 6) μrad in phase displacement between the CVD component and system calibrations up to 100 kV. Requirements for achieving this agreement are adequate grounding of the CVD and the use of triax cable between the high-voltage (HV) capacitor and the CVD low-voltage electronics in case large distances have to be covered on-site. The (4±6) μrad agreement in phase displacement is well within the required 10 μrad limit for voltage measurements as part of on-site TLM system calibrations with 20 μW/VA overall uncertainty at low power factors.
ISSN:0018-9456
1557-9662
DOI:10.1109/TIM.2021.3056647