Determination of ethylene by field asymmetric ion mobility spectrometry
The determination of ethylene with a field asymmetric ion mobility spectrometer, which can easily be constructed in-house, is described. The device makes use of a Krypton lamp for ionization. A rectangular pulse of 500 V pp at 1 MHz was employed as separation waveform in the drift tube rather than t...
Gespeichert in:
Veröffentlicht in: | International journal for ion mobility spectrometry 2020-10, Vol.23 (2), p.161-166 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The determination of ethylene with a field asymmetric ion mobility spectrometer, which can easily be constructed in-house, is described. The device makes use of a Krypton lamp for ionization. A rectangular pulse of 500 V
pp
at 1 MHz was employed as separation waveform in the drift tube rather than the commonly used less efficient bisinusoidal waveform. The calibration curve for the range from 670 ppb
(
V
/V)
to 67 ppm
(V/V)
was found to be highly linear with a correlation coefficient of r = 0.9999. The limit of detection was determined as 200 ppb
(
V
/V)
. The reproducibility was 4% (relative standard deviation). The device was found to be suitable for the determination of ethylene given off by fruit; 6 types of climacteric fruit were tested, namely apples, bananas, kiwi fruit, nectarines, pears and plums. |
---|---|