CALIBRATION OF THE EXTENDED QUADRIHEDRAL NaI(Tl) SPECTROMETER BY COSMIC MUONS

The results of cosmic muon calibration of the extended quadrihedral NaI(Tl) spectrometer are presented. The signal amplitude and energy resolution of the spectrometer in the direction transverse to the spectrometer axis have complex dependences on the photomultiplier voltage and particle entrance po...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Bulletin of the Lebedev Physics Institute 2020-08, Vol.47 (8), p.219-222
Hauptverfasser: Baskov, V. A., Govorkov, B. B., Polyansky, V. V.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The results of cosmic muon calibration of the extended quadrihedral NaI(Tl) spectrometer are presented. The signal amplitude and energy resolution of the spectrometer in the direction transverse to the spectrometer axis have complex dependences on the photomultiplier voltage and particle entrance point to the spectrometer. At a photomultiplier divider voltage U = 1150 ± 10 V, the region of uniform signals is observed over ~90% of the spectrometer length, the relative energy resolution is δ ≈ 7%.