The Effect of Different Irrigation Solutions on the Accuracy of Two Electronic Apex Locators in Locating Artificial Root Perforations
Objective: This study aimed to assess the accuracy of the Raypex 6 (VDW GmbH, Munich, Germany) and Root ZX mini (J. Morita Corp., Tokyo, Japan) electronic apex locators (EALs) with regard to detecting root perforations in dry conditions and in the presence of irrigation solutions [2.5% sodium hypoch...
Gespeichert in:
Veröffentlicht in: | Meandros medical and dental journal 2020-08, Vol.21 (2), p.134-139 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective: This study aimed to assess the accuracy of the Raypex 6 (VDW GmbH,
Munich, Germany) and Root ZX mini (J. Morita Corp., Tokyo, Japan) electronic apex
locators (EALs) with regard to detecting root perforations in dry conditions and
in the presence of irrigation solutions [2.5% sodium hypochlorite (NaOCl), 17%
ethylenediaminetetraacetic acid (EDTA) and Qmix].
Materials and Methods: Twenty single-rooted human teeth were perforated
artificially with a 1 mm diameter in the middle region. The actual canal lengths
up to the perforation site were determined, and then the teeth were embedded in
an alginate mold. The electronic measurements of the perforations were obtained
using a size #20 K-file for each EAL in various conditions.
Results: There were significant differences between the different canal conditions
in the Raypex 6 group (p0.05). There was a statistically significant difference (p0.05).
Conclusion: Under the conditions of this study, both apex locators detected
root canal perforations within a clinically acceptable (range of -0.18 to 0.31mm)
distance from the coronal border of the perforation region. Irrigation solutions
within the root canal affected the accuracy of Raypex 6, but not of Root ZX mini. |
---|---|
ISSN: | 2149-9063 2149-9063 |
DOI: | 10.4274/meandros.galenos.2020.41196 |