An Efficient HPLC–MS Method for Impurity Profile of Ertapenem
A mass spectrometer compatible, reverse-phase high-performance liquid chromatography method capable to resolve twenty-six impurities and degradation products (DPs) in Ertapenem (ERT) has been developed. Eight dimer impurities ( m / z 951) and two dehydrated dimers ( m / z 933) were baseline resolved...
Gespeichert in:
Veröffentlicht in: | Chromatographia 2020-09, Vol.83 (9), p.1095-1105 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A mass spectrometer compatible, reverse-phase high-performance liquid chromatography method capable to resolve twenty-six impurities and degradation products (DPs) in Ertapenem (ERT) has been developed. Eight dimer impurities (
m
/
z
951) and two dehydrated dimers (
m
/
z
933) were baseline resolved. Out of eight dimers, four dimers are reported the first time. The impurities and DPs were identified using LC–MS/MS. The chromatographic separation was achieved on X-Terra RP 18 (150 × 4.6 mm, 3.5 µm) column using gradient elution. The separation was dependent on the pH of the mobile phase, optimum separation was achieved using the mobile phase of pH 8. Mobile phase-A comprised of a mixture of ammonium formate buffer (pH 8.0), water and acetonitrile in a ratio of 40:60:01; v/v/v. Mobile phase-B comprised of a mixture of ammonium formate buffer (pH 8.0), acetonitrile and methanol in a ratio of 40:50:10; v/v/v. The flow rate was 1.0 mL min
−1
and column was maintained at 40 °C. Considering the fragility of ERT and demand of faster analysis method, the method was further transformed on rapid resolution column with a run time of 10 min. The later method also produced the same kind of selectivity with enhanced sensitivity. |
---|---|
ISSN: | 0009-5893 1612-1112 |
DOI: | 10.1007/s10337-020-03928-6 |