POCIS Calibration for Organic Compound Sampling in Small Headwater Streams

Field‐based atrazine sampling rates (Rs) obtained by the polar organic chemical integrative sampler (POCIS) method were measured in 9 headwater streams over 3 yr covering 5 to 6 exposure periods of 2 to 3 wk/site/yr. Rates were best in line with the model Rs = 148 mL/d, with a standard deviation of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Environmental toxicology and chemistry 2020-07, Vol.39 (7), p.1334-1342
Hauptverfasser: Booij, Kees, Chen, Sunmao, Trask, Jennifer R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Field‐based atrazine sampling rates (Rs) obtained by the polar organic chemical integrative sampler (POCIS) method were measured in 9 headwater streams over 3 yr covering 5 to 6 exposure periods of 2 to 3 wk/site/yr. Rates were best in line with the model Rs = 148 mL/d, with a standard deviation of 0.17 log units (factor 1.5). The POCIS canisters reduced mass transfer coefficients of the water boundary layer by a factor of 2 as measured by alabaster dissolution rates. A mechanistic model that accounts for flow and temperature effects yielded a fair estimate of the effective exchange surface area (12.5 ± 0.8 cm2). This model could only be tested for higher flow velocities because of uncertainties associated with the measurement of flow velocities
ISSN:0730-7268
1552-8618
DOI:10.1002/etc.4731