On-chip estimation of hematocrit level for diagnosing anemic conditions by Impedimetric techniques
An on-chip device has been fabricated on Si platform for precise estimation of the hematocrit (Hct) level of human blood with rapid turn out. Impedance/capacitance spectroscopy and current-voltage (I-V) measurements have been employed to observe the variation of electrical parameters of erythrocyte...
Gespeichert in:
Veröffentlicht in: | Biomedical microdevices 2020-06, Vol.22 (2), p.38-38, Article 38 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An on-chip device has been fabricated on Si platform for precise estimation of the hematocrit (Hct) level of human blood with rapid turn out. Impedance/capacitance spectroscopy and current-voltage (I-V) measurements have been employed to observe the variation of electrical parameters of erythrocyte suspensions with varying Hct level. Experimentally obtained values of capacitance, impedance and conductance with Hct level suggests a linear variation. Current-time measurement has also been performed to ensure the susceptibility of red blood cells under a fixed external electric field for certain duration of time. The online real time sample analyzes have also been performed by the device connected with an embedded electronic circuit interfaced with a laptop through appropriate software. This has enabled the development of a novel Si-chip based compact point-of-care (POC) diagnosis system for Hct level estimation by measuring the dielectric/capacitive variation of an erythrocyte cell suspension. The relevant performance parameters of such a compact system including range, resolution, limit of detection and throughput have also been evaluated. |
---|---|
ISSN: | 1387-2176 1572-8781 |
DOI: | 10.1007/s10544-020-00493-5 |