Dose distribution verification in intraoperative radiation therapy using an N-isopropyl acrylamide-based polymer gel dosimeter
The purpose of the current study was to verify the dose distribution of an Intrabeam-50 kV IORT system using polymer gel dosimetry technique. Results of dose distribution evaluation using NIPAM polymer gel dosimetry were compared with those measured using an ionization chamber and simulated using MC...
Gespeichert in:
Veröffentlicht in: | Journal of radioanalytical and nuclear chemistry 2020-05, Vol.324 (2), p.481-488 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of the current study was to verify the dose distribution of an Intrabeam-50 kV IORT system using polymer gel dosimetry technique. Results of dose distribution evaluation using NIPAM polymer gel dosimetry were compared with those measured using an ionization chamber and simulated using MCNPX code. Results showed the calculated gamma index was less than 1 with 2% dose-difference/2 mm distance-to-agreement for comparison between NIPAM and ionization chamber as well as between NIPAM and MCNPX simulation. It was concluded that the NIPAM polymer gel dosimetry is useful for verifying the dose distribution of low energy X-ray IORT technique. |
---|---|
ISSN: | 0236-5731 1588-2780 |
DOI: | 10.1007/s10967-020-07110-y |