Calibration of Photoelectric Pulsed Radiation Detectors
A method for calibrating a photodiode intended for high-intensity radiation pulse detection using continuous lighting lamps is described. The method is based on the use of an attachable integrated circuit with a known time constant. Based on the volt sensitivity and the time constant obtained during...
Gespeichert in:
Veröffentlicht in: | Russian physics journal 2020-03, Vol.62 (11), p.2046-2051 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A method for calibrating a photodiode intended for high-intensity radiation pulse detection using continuous lighting lamps is described. The method is based on the use of an attachable integrated circuit with a known time constant. Based on the volt sensitivity and the time constant obtained during calibration against a continuous lighting lamp, relationships of the power and radiation energy with the photodetector response are mathematically derived. The method has been tested using a xenon flash lamp with pulse duration at half maximum of ≈60 μs. |
---|---|
ISSN: | 1064-8887 1573-9228 |
DOI: | 10.1007/s11182-020-01943-6 |