Calibration of Photoelectric Pulsed Radiation Detectors

A method for calibrating a photodiode intended for high-intensity radiation pulse detection using continuous lighting lamps is described. The method is based on the use of an attachable integrated circuit with a known time constant. Based on the volt sensitivity and the time constant obtained during...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Russian physics journal 2020-03, Vol.62 (11), p.2046-2051
Hauptverfasser: Kireev, S. G., Tumashevich, K. A., Shashkovskiy, S. G., Abakumov, A. V., Osin, N. A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A method for calibrating a photodiode intended for high-intensity radiation pulse detection using continuous lighting lamps is described. The method is based on the use of an attachable integrated circuit with a known time constant. Based on the volt sensitivity and the time constant obtained during calibration against a continuous lighting lamp, relationships of the power and radiation energy with the photodetector response are mathematically derived. The method has been tested using a xenon flash lamp with pulse duration at half maximum of ≈60 μs.
ISSN:1064-8887
1573-9228
DOI:10.1007/s11182-020-01943-6