X-ray exposure in odontoid screwing for Anderson type II fracture: comparison between O-arm and C-arm-assisted procedures
Background Since the odontoid fractures become increasingly common in the aging population, technical improvements are even more needed. The odontoid screwing has been progressively preferred by many surgeons in type II fractures according to the Anderson-D’Alonzo classification system. However, X-r...
Gespeichert in:
Veröffentlicht in: | Acta neurochirurgica 2020-03, Vol.162 (3), p.713-718 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
Since the odontoid fractures become increasingly common in the aging population, technical improvements are even more needed. The odontoid screwing has been progressively preferred by many surgeons in type II fractures according to the Anderson-D’Alonzo classification system. However, X-ray exposure remains an issue for surgeons and OR staff members. The aim of the present study was to investigate the feasibility of using the O-Arm for odontoid screwing comparing the radiation exposure to the standard C-Arm.
Methods
Patients consequently referred to our center for odontoid type II fractures, from January 2018 to April 2019, eligible for odontoid screwing were enrolled in the present study. They were operated on using either C-Arm or O-Arm-assisted procedures. The surgical duration, number of acquisitions, global X-ray exposure for the OR staff and patients, and screw placement accuracy were evaluated.
Results
No differences in terms of patients’ demographical characteristics and surgical duration were reported. The number of acquisitions, intraoperative and global X-ray dose, for the OR staff and patients, was lower in O-Arm-assisted procedures (
p
< 0.05). The screws were all well positioned.
Conclusions
Since the surgical outcomes seem to be similar using the O-Arm for odontoid screwing, the lower X-ray exposure and the possibility for checking the instrumentation positioning with 3D reconstructions before leaving the OR should be considered. |
---|---|
ISSN: | 0001-6268 0942-0940 |