Measurement of the time resolution of small SiPM-based scintillation counters

In this research, we evaluated the timing resolution of SiPM-based scintillation detector on a 1-GeV electron beam “extracted” from VEPP-4M. We tested small scintillation crystals of pure CsI, YAP, LYSO, and LFS-3 with HAMAMATSU S10362-33-025C and S13360-3050CS. The CsI scintillator together with HA...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of instrumentation 2017-12, Vol.12 (12), p.P12020-P12020
Hauptverfasser: Kravchenko, E.A., Porosev, V.V., Savinov, G.A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this research, we evaluated the timing resolution of SiPM-based scintillation detector on a 1-GeV electron beam “extracted” from VEPP-4M. We tested small scintillation crystals of pure CsI, YAP, LYSO, and LFS-3 with HAMAMATSU S10362-33-025C and S13360-3050CS. The CsI scintillator together with HAMAMATSU S13360-3050CS demonstrated the best results. Nevertheless, the achieved time resolution of ∼80 ps (RMS) relates mainly to the photodetector itself. It makes the silicon photomultiplier an attractive candidate to replace other devices in applications where sub-nanosecond accuracy is required.
ISSN:1748-0221
1748-0221
DOI:10.1088/1748-0221/12/12/P12020