Update of national diagnostic reference levels for adult CT in Switzerland and assessment of radiation dose reduction since 2010
Objectives To update the national diagnostic reference levels (DRLs) for adult CT in Switzerland using dose management software and to compare them to the previous Swiss DRLs from 2010. Methods CT dose data from 14 radiological institutes with a total of 50 CT scanners were collected with locally in...
Gespeichert in:
Veröffentlicht in: | European radiology 2020-03, Vol.30 (3), p.1690-1700 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objectives
To update the national diagnostic reference levels (DRLs) for adult CT in Switzerland using dose management software and to compare them to the previous Swiss DRLs from 2010.
Methods
CT dose data from 14 radiological institutes with a total of 50 CT scanners were collected with locally installed dose management software between 2014 and 2017. Data were assigned to 15 defined master protocols. Data cleaning steps were developed and adjusted individually for each participating institute and protocol. The DRLs for each master protocol were calculated as the 75th percentile of the distribution of the median volume computed tomography dose index (CTDI
vol
) and dose-length product (DLP) values per CT scanner.
Results
In total, 220,269 CT exams were available after data cleaning. Updated DRLs showed a clear trend towards lower doses compared with previous DRLs. The average relative change in the DRLs for CTDI
vol
was − 30% (0 to − 47%) and − 22% for DLP (+ 20 to − 40%). The largest relative decrease in the DRL for DLP was observed for the cervical spine protocol (− 40%), the two chest protocols (chest, − 37%; and exclusion of pulmonary embolism, − 33%), and the two neck protocols (neck, − 32%; and carotid angiography, − 28%). The DRLs for other protocols, for example the head and the abdomen-pelvis protocol, showed smaller relative changes (− 11% and − 17%).
Conclusions
The updated national DRLs are substantially lower than the previous values from 2010, demonstrating technological progress and the efforts of the radiological community to lower CT radiation exposure.
Key Points
• Dose management software allows the establishment of DRLs based on big data.
• Updated Swiss DRLs for adult CT are substantially lower compared with those from 2010.
• Swiss DRLs are low compared with other national DRLs. |
---|---|