Transfemoral transcatheter aortic valve implantation after aortic valve repair with HAART 300 device

We report the first successful case, to our knowledge, of CoreValve Evolut R (Medtronic, Minneapolis, MN) implantation into a failed HAART 300 aortic annuloplasty device (BioStable Science & Engineering, TX). An 81‐year‐old man presented with severe symptomatic aortic regurgitation secondary to...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Catheterization and cardiovascular interventions 2019-11, Vol.94 (6), p.856-858
Hauptverfasser: Takayuki, Gyoten, Volker, Herwig, Axel, Harnath, Dirk, Fritzsche
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report the first successful case, to our knowledge, of CoreValve Evolut R (Medtronic, Minneapolis, MN) implantation into a failed HAART 300 aortic annuloplasty device (BioStable Science & Engineering, TX). An 81‐year‐old man presented with severe symptomatic aortic regurgitation secondary to failure of the 21 mm HAART 300 device, which had been implanted 45 days previously. Transthoracic echocardiography (TTE) revealed grade 3 aortic regurgitation with central jet, without aortic valve stenosis. Because of the high risk for redo surgery, the heart team proceeded with femoral transcatheter aortic valve implantation. The 26 mm CoreValve Evolut R was deployed into the 21 mm HAART 300 device without difficulty or complications. There were no intraoperative or postoperative complications. The patient was discharged after 5 days. TTE showed a mean aortic valve gradient of 18 mmHg, with minimal paravalvular leak. Our experience suggests that CoreValve Evolut R implantation may be an attractive option in patients with failed HAART 300 aortic annuloplasty.
ISSN:1522-1946
1522-726X