Letters to the Editor Comment on The Radiological Physics Center's standard dataset for small field size output factors (J Appl Clin Med Phys. 2012;13(5):282‐89)
The measurements were performed on Varian TrueBeam accelerators (Varian Medical Systems, Palo Alto, CA) using three different detectors all suited for small field dosimetry: a pinpoint chamber (PTW 31014; PTW, Freiburg, Germany), a prototype scintillator system developed at Technical University of D...
Gespeichert in:
Veröffentlicht in: | Journal of applied clinical medical physics 2014-03, Vol.15 (2), p.350-352 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The measurements were performed on Varian TrueBeam accelerators (Varian Medical Systems, Palo Alto, CA) using three different detectors all suited for small field dosimetry: a pinpoint chamber (PTW 31014; PTW, Freiburg, Germany), a prototype scintillator system developed at Technical University of Denmark, (2) and a diamond detector (PTW 60003). The measured RPC values were compared to the measured values at the participating centers with site, field size, detector type, and energy as variables. For 6 MV, the ANOVA estimated the output factors at the Danish centers to be larger than RPC data by a value of 0.0103 (about 1.2%), with a 95% confidence interval of [0.0084; 0.0123]. |
---|---|