Demonstration of a New Technique Using Radioisotope Tracers to Measure the Backside Wear Rate on Tibial Inserts

Local backside wear measurements on ultra-high molecular weight polyethylene (UHMWPE) tibial inserts in LCS mobile bearing knee prostheses have been performed using a new radioisotope tracing technique. The radioisotope tracers 97 Ru and 101m Rh were synthesized via a fusion evaporation reaction and...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Tribology letters 2012-05, Vol.46 (2), p.139-145
Hauptverfasser: Warner, Jacob A., Gladkis, Laura G., Smith, Paul N., Scarvell, Jennifer M., Timmers, Heiko
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Local backside wear measurements on ultra-high molecular weight polyethylene (UHMWPE) tibial inserts in LCS mobile bearing knee prostheses have been performed using a new radioisotope tracing technique. The radioisotope tracers 97 Ru and 101m Rh were synthesized via a fusion evaporation reaction and recoil-implanted into cylindrical plugs of UHMWPE. The labelled plugs were carefully fitted into tibial inserts at two relevant locations. With bovine serum acting as a lubricant, the tibial inserts were then worn in vitro for 500,000 and 780,000 cycles, respectively, in a pneumatic knee motion simulator. Results reflect the non-linear change of wear during the wear-in phase and its evolution to a long-term steady-state rate. This new technique shows potential for extracting localized wear rates across the backside of a tibial insert in order to develop a comprehensive backside wear model.
ISSN:1023-8883
1573-2711