Dorsal screw penetration following implant removal after volar locked plating of distal radius fracture
Complications while removing implants are quite common. In rare cases screw shanks must be left in situ. We present a case of a 38-year-old patient who was treated using a palmar locking plate for a distal radius fracture. After implant removal, one of the screw shanks left in situ penetrated the do...
Gespeichert in:
Veröffentlicht in: | Archives of orthopaedic and trauma surgery 2011-09, Vol.131 (9), p.1279-1282 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Complications while removing implants are quite common. In rare cases screw shanks must be left in situ. We present a case of a 38-year-old patient who was treated using a palmar locking plate for a distal radius fracture. After implant removal, one of the screw shanks left in situ penetrated the dorsal cortex of the distal radius into the third extensor tendon compartment and led to irritation of the extensor pollicis longus tendon. We report the unexpected complication and analysis of possible errors of this case to avoid this kind of complication. |
---|---|
ISSN: | 0936-8051 1434-3916 |
DOI: | 10.1007/s00402-011-1300-y |