Spurious hyperphosphatemia in a patient with alteplase-locked central venous catheter

Alteplase has been shown to be effective in preventing central venous access clotting in patients on hemodialysis. Because of a high phosphorus content in its excipient, it can inadvertently contaminate blood samples, leading the physician in care of the patient to erroneously increase dialysis time...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pediatric nephrology (Berlin, West) West), 2006-02, Vol.21 (2), p.301-302
Hauptverfasser: Cachat, Francois, Bardy, Daniel, Durussel, Claudine, Di Paolo, Ermindo
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Alteplase has been shown to be effective in preventing central venous access clotting in patients on hemodialysis. Because of a high phosphorus content in its excipient, it can inadvertently contaminate blood samples, leading the physician in care of the patient to erroneously increase dialysis time or change diet in order to control the pseudo-hyperphosphatemia.
ISSN:0931-041X
1432-198X
DOI:10.1007/s00467-005-2088-1