Validity of pulse oximetry during maximal exercise in normoxia, hypoxia, and hyperoxia
Department of Medicine, University of California, San Diego, La Jolla, California 92093 During exercise, pulse oximetry is problematic due to motion artifact and altered digital perfusion. New pulse oximeter technology addresses these issues and may offer improved performance. We simultaneously comp...
Gespeichert in:
Veröffentlicht in: | Journal of applied physiology (1985) 2002-01, Vol.92 (1), p.162-168 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Department of Medicine, University of California, San Diego, La
Jolla, California 92093
During exercise, pulse
oximetry is problematic due to motion artifact and altered digital
perfusion. New pulse oximeter technology addresses these issues and may
offer improved performance. We simultaneously compared Nellcor N-395
(Oxismart XLTM) pulse oximeters with an RS-10 forehead sensor (RS-10),
a D-25 digit sensor (D-25), and the Ivy 2000 (Masimo SETTM)/LNOP-Adt
digit sensor (Ivy) to arterial blood oxygen saturation
(Sa O 2 ) by cooximetry. Nine normal subjects, six
athletes, and four patients with chronic disease exercised to maximum
oxygen consumption ( O 2 max ) under various conditions [normoxia, hypoxia inspired oxygen fraction (F I O 2 ) = 0.125; hyperoxia,
F I O 2 = 1.0]. Regression analysis for normoxia and hypoxic data was performed ( n = 161 observations, Sa O 2 = 73-99.9%), and bias
(B) and precision (P) were calculated. RS10 offered greater validity
than the other two devices tested ( y = 1.009 x 0.52, R 2 = 0.90, B±P = 0.3 ± 2.5). Finger sensors had low precision and a
significant negative bias (D-25: y = 1.004 x 2.327, R 2 = 0.52, B±P = 2.0 ± 7.3; Ivy: y = 1.237 x 24.2, R 2 = 0.78, B±P = 2.0 ± 5.2). Eliminating measurements in which heart
rate differed by >10 beats/min from the electrocardiogram value
improved precision minimally and did not affect bias substantially (B±P = 0.5 ± 2.0, 1.8 ± 8.4, and 1.25±4.33 for
RS-10, D-25, and Ivy, respectively). Signal detection algorithms and
pulse oximeter were identical between RS-10 and D-25; thus the improved performance of the forehead sensor is likely because of sensor location. RS-10 should be considered for exercise testing in which pulse oximetry is desirable.
oxygen saturation; cooximetry; patients; normal subjects; athletes |
---|---|
ISSN: | 8750-7587 1522-1601 |
DOI: | 10.1152/japplphysiol.00409.2001 |