Elucidation of the statistical factors that influence anti-drug antibody cut point setting through a multi-laboratory study
Appropriateness of anti-drug antibody (ADA) assay is critical for immunogenicity assessment of biopharmaceuticals. Although cut point setting in ADA assay has a large impact on the results, a standard statistical approach for its setting has not been well established. In this multi-laboratory study,...
Gespeichert in:
Veröffentlicht in: | Bioanalysis 2019-03, Vol.11 (6), p.509-524 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Appropriateness of anti-drug antibody (ADA) assay is critical for immunogenicity assessment of biopharmaceuticals. Although cut point setting in ADA assay has a large impact on the results, a standard statistical approach for its setting has not been well established.
In this multi-laboratory study, to elucidate factors influencing the cut point setting, we compared the statistical approaches and calculated cut points for multiple datasets of ADA assays using the individual procedure employed at each laboratory.
We showed that outlier exclusion, false-positive rate and investigating data distribution have the greatest impact on both screening and confirmatory cut points. Our results would be useful for industry researchers and regulators engaged in immunogenicity assessment of biopharmaceuticals. |
---|---|
ISSN: | 1757-6180 1757-6199 |
DOI: | 10.4155/bio-2018-0178 |