Dosimetric characterization of a Cesium-137 irradiator for irradiation of cell cultures

In the Instituto Nacional de Cancerología (INCan), new treatments involving the use of radiotherapy are investigated, evaluating their effects on cell cultures. For irradiation it is necessary to characterize a 137Cs BIOBEAM-GM 2000 irradiator. The characterization was carried out using a dosimetric...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Hernández, Manuel Silva, Víquez, Itzel Militza Torres, Medina, Luis A.
Format: Tagungsbericht
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In the Instituto Nacional de Cancerología (INCan), new treatments involving the use of radiotherapy are investigated, evaluating their effects on cell cultures. For irradiation it is necessary to characterize a 137Cs BIOBEAM-GM 2000 irradiator. The characterization was carried out using a dosimetric system composed of a radiochromic dye film GAFCHROMIC R EBT3, a Scanmaker 8700 Microteck scanner and the Matlab R2016b program. As a result of the characterization, a dose rate of 4.8 ±0.4 Gy min−1 and a distribution with a radial dose gradient over three cell culture dishes are obtained. In conclusion, a protocol was established to irradiate three cell culture dishes at the same dose in a single exposure.
ISSN:0094-243X
1551-7616
DOI:10.1063/1.5095895