Study of Hydrogen Peroxide Vapors Sensors Based on Carbon Nanotubes Coated with Tin Oxide Nanoparticles
We present the results of studies of the nanocomposite MWCNTs/SnO 2 hydrogen peroxide vapor sensors. The technology of manufacturing of these sensors has been developed. Because of the measurements of the temperature characteristics, the 100°C optimal operating temperature of the studied sensors has...
Gespeichert in:
Veröffentlicht in: | Journal of contemporary physics 2019, Vol.54 (1), p.57-64 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present the results of studies of the nanocomposite MWCNTs/SnO
2
hydrogen peroxide vapor sensors. The technology of manufacturing of these sensors has been developed. Because of the measurements of the temperature characteristics, the 100°C optimal operating temperature of the studied sensors has been found. The response and recovery curves of the sensors were investigated in the presence of different concentrations of hydrogen peroxide vapor in the atmosphere. A sufficiently high response was registered already at close to the threshold low concentrations of the target gas present in the air. In a double logarithmic scale, the linear dependence of the sensors response on the concentration of hydrogen peroxide vapor was observed in a certain concentration range. The minimal registered gas concentration is less than 1 ppm. |
---|---|
ISSN: | 1068-3372 1934-9378 |
DOI: | 10.3103/S1068337219010079 |