A New Load-based Thermopower Measurement Setup in the Temperature Range of 5 - 330 K
This is the report on new development of an automated precision load-based measurement setup for thermoelectric power of different types of samples in the temperature range 5 - 330 K. The problems in the old spring-based setup have been solved by making load-based setup. This new setup takes nearly...
Gespeichert in:
Veröffentlicht in: | arXiv.org 2019-03 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This is the report on new development of an automated precision load-based measurement setup for thermoelectric power of different types of samples in the temperature range 5 - 330 K. The problems in the old spring-based setup have been solved by making load-based setup. This new setup takes nearly 4 h for each run and typical error is within 5 %. A high quality calibration been demonstrated using high purity platinum wires. |
---|---|